Descripción

Measurement Head: High-precision model
Model IM-6145
Type Measurement head
Image sensor 1' 6.6 mega pixel CMOS
Display 10.4' LCD monitor (XGA: 1024 ? 768) External monitor connectable (clone output)
Receiver lens Double telecentric lens
Image measurement Field of view Wide-field measurement mode: o25 mm High-precision measurement mode: o6 mm
Minimum display unit 0.1 µm
Repetition accuracy Wide-field measurement mode: ±0.5 µm High-precision measurement mode: ±0.1 µm
Measurement accuracy (±2?) Wide-field measurement mode: ±2 µm*1 High-precision measurement mode: ±0.7 µm*2
External remote input No-voltage input (with contacts, without contacts)
External output OK/NG/FAIL/MEAS. Relay outputRated load 24 VDC 0.5AON resistance 50 m? or less
Interface LAN RJ-45 (10BASE-T / 100BASE-TX / 1000BASE-T)
USB 2.0 series A 6 ports (front: 2, rear: 4)
Record Hard disk drive 250 GB
Illumination system Transparent Telecentric transparent illumination (Green LED)
Ring Four division ring illumination (White LED)
Coaxial epi-illumination (white LED)
External Externally connected illumination
X stage Moving range Within o25 mm field
Z stage 20 mm
Withstand load 1 kg
Y stage Moving range Within o25 mm field
Power supply Power voltage 100 to 240 VAC 50 / 60 Hz
Power consumption 310 mA or less
Environmental resistance Ambient temperature +10 to +35 °C
Relative humidity 20 to 80 % RH (No condensation)
Weight Approx 25 kg
*1 In the range of o20 mm from the centre of the stage at the operating temperature range of +23 ±1°C at the focused focal point position *2 In the range of o5 mm from the centre of the stage at the operating temperature range of +23 ±1°C at the focused focal point position
''